Prediction of flow capability in intravenous infusion systems: implications for fluid resuscitation
1Department of Anesthesia, Brigham and Women's Hospital, Harvard Medical School, Boston, MA 02115.
Abstract:
The pressure-flow (P-F) relationship for intravenous infusion systems is nonlinear and may be expressed by the quadratic model P = RLF + RTF2. The flow parameters RL and RT may represent the resistance of laminar and turbulent flow, respectively. In this study pressure and flow were measured, and RL and RT were calculated for several infusion tubings, catheter, and system components. We then developed a technique to identify the relative effect of various devices on achieving the higher flows needed for fluid resuscitation. A typical infusion system was chosen, and the experimentally determined flow parameters RL and RT of its components were used in the quadratic P-F relationship at P = 300 mm Hg. Devices in the infusion system were ranked, using a subtractive algorithm, according to their relative impediment to flow as measured by contribution to the pressure drop. The order of devices removed or replaced, from largest to least pressure drop, was as follows: fluid warmer, 16-gauge catheter, check valve, 14-gauge catheter, standard-bore Y tubing, 12-gauge catheter, and standard-bore stopcock, leaving 10-gauge catheter + wide tubing. Devices with large RT, such as fluid warmers and check valves that produce large pressure drops, should generally be avoided during fluid resuscitation when high flows are needed. A similar ordering of device removal or substitution (largest to least pressure drop) was determined using the traditional but incorrect linear P-F model, P = RF, and the order of devices chosen for elimination was different.(ABSTRACT TRUNCATED AT 250 WORDS)
Related Concept Videos
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations