Related Experiment Video
Updated: May 13, 2026

Designing a Bioreactor to Improve Data Acquisition and Model Throughput of Engineered Cardiac Tissues
Published on: June 2, 2023
Reinforcement of HeartMate II bend relief connection: champagne bottle technique
Takeyoshi Ota1, Hiroo Takayama, Zain Khalpey
1Division of Cardiothoracic Surgery, Department of Surgery, Columbia University, New York, New York 10032, USA. to2226@columbia.edu
Abstract:
HeartMate II (Thoratec Corp, Pleasanton, CA) continuous-flow ventricular assist devices may be exchanged through a subcostal approach, where the device body can be accessed easily and safely. The existing bend relief needs to be reattached to the new device body at the end of the operation, and this reattachment often results in a malalignment between the outflow graft, the bend relief, and the device body. This causes unexpected detachment and subsequent graft kinking. We present a new, simple technique to reinforce the bend relief connection.