[New device and technique to measure intra-abdominal pressure]
Jorge Alberto Castañón-González1, Julián Satué-Rodríguez, Francisco Carrillo Rosales
1Universidad Anáhuac, Huixquilucan, Estado de México, México. jorgec@prodigy.net.mx
Introduction:
sequential measurement of intra-abdominal pressure is of paramount importance for an early detection and appropriate therapeutic management of intra-abdominal hypertension and abdominal compartment syndrome.
Objective:
to validate a device and technique developed to measure intra-abdominal pressure (an innovation of the Foley urinary catheter named intra-abdominal pressure catheter).
Methods:
three different sets of measurements where done to test the intra-abdominal pressure catheter device: I. 50 measurements were done with the intra-abdominal pressure catheter device and compared against those measurements done using a manometer graded in cm H(2)O; II. Direct measurement of intra-abdominal pressure in five patients during elective laparoscopy vs the intraabdominal pressure catheter device; and III. Measurement of intra-abdominal pressure by the Kron method (Gold standard) vs intra-abdominal pressure catheter device in three patents with intra-abdominal hypertension/abdominal compartment syndrome. Measurements where compared with Pearson correlation test and Bland Altman statistics.
Results:
I.Intra-abdominal pressure catheter vs graded manometer: r = 0.99, with a mean pressure difference of 0.27 ± 0.23 mmHg, CI (0.039 to 0.092 mmHg). II. Direct measurement of intra-abdominal pressure during laparoscopy vs intra-abdominal pressure catheter device: r = 0.93, with a mean pressure difference of 0.18 ± 0.84 mmHg, CI (-1.46 to 1.83 mmHg) and III. Measurement of intra-abdominal pressure by the Kron Method vs intra-abdominal pressure catheter device: r = 0.81, with a mean pressure difference of -0.41 ± 0.87, CI (-2.12 mmHg to 1.30 mmHg).
Conclusions:
the intra-abdominal pressure catheter device is a safe and reliable instrument for measuring intra-abdominal pressure.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of the Abdomen III: Palpation
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement