Anterior chamber angle imaging with swept-source optical coherence tomography: measuring peripheral anterior synechia
Isabel Lai1, Heather Mak, Gilda Lai
1Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong, PR China.
Objective:
To investigate the use of swept-source optical coherence tomography (OCT) for measuring the area and degree of peripheral anterior synechia (PAS) involvement in patients with angle-closure glaucoma.
Design:
Cross-sectional study.
Participants:
Twenty-three eyes with PAS (detected by indentation gonioscopy) from 20 patients with angle-closure glaucoma (20 eyes had primary angle-closure glaucoma and 3 eyes had angle-closure glaucoma secondary to chronic anterior uveitis [n = 2] and Axenfeld-Rieger syndrome [n = 1]).
Methods:
The anterior chamber angles were evaluated with indentation gonioscopy and imaged by swept-source OCT (Casia OCT, Tomey, Nagoya, Japan) in room light and in the dark using the "angle analysis" protocol, which was composed of 128 radial B-scans each with 512 A-scans (16-mm scan length). The area and degree of PAS involvement were measured in each eye after manual detection of the scleral spur and the anterior irido-angle adhesion by 2 masked observers. The interobserver variability of the PAS measurements was calculated.
Main Outcome Measures:
The agreement of PAS assessment by gonioscopy and OCT, the area and the degree of PAS involvement, and the intraclass correlation coefficient (ICC) of interobserver PAS measurements.
Results:
The area of PAS (mean ± standard deviation) was 20.8 ± 16.9 mm(2) (range, 3.9-74.9 mm(2)), and the degree of PAS involvement was 186.5 ± 79.9 degrees (range, 42-314 degrees). There was no difference in the area of PAS (P = 0.90) and the degree of PAS involvement (P = 0.95) between images obtained in room light and in the dark. The interobserver ICCs were 0.99 (95% confidence interval [CI], 0.98-1.00) for the area of PAS and 0.99 (95% CI, 0.97-1.00) for the degree of PAS involvement. There was good agreement of PAS assessment between gonioscopy and OCT images (kappa = 0.79; 95% CI, 0.67-0.91).
Conclusions:
Swept-source OCT allows visualization and reproducible measurements of the area and degree of PAS involvement, providing a new paradigm for evaluation of PAS progression and risk assessment for development of angle-closure glaucoma.
Financial Disclosure(S):
The author(s) have no proprietary or commercial interest in any materials discussed in this article.