Related Experiment Video
Updated: May 12, 2026

Implantation and Control of Wireless, Battery-free Systems for Peripheral Nerve Interfacing
Published on: October 20, 2021
S3 motor branch stimulation failure due to nerve fiber burning at the nerve-wire junction: A historical technical
Zahra Emami1, Najme-Sadat Moosavi, Majid Karimi
1Department of Neurosurgery, Sina Trauma and Surgery Research Center, Tehran University of Medical Sciences, Tehran, Iran.
Background:
Sacral nerve stimulation is a minimally invasive procedure to treat spinal cord injured (SCI) patients with overactive bladder syndrome or nonobstructive urinary retention that is refractory to conservative treatment.
Methods:
In this paper, we report a case of traumatic cervical SCI with quadriplegia and spastic bladder, which was managed by third sacral motor branch stimulation in 1998.
Results:
In this case, stimulation-induced burning of nerve fibers was seen microscopically during the implantation surgery. At 2 weeks after the index surgery, the stimulator was removed due to ineffectiveness. We hypothesize that the stimulation settings of our stimulator were not appropriate for neural stimulation and led to neural destruction, fibrosis, and treatment failure.
Conclusion:
The device settings of stimulators used in neural stimulation should be appropriate for direct neural stimulation otherwise they can lead to neural destruction and treatment failure.