Related Experiment Video
Updated: May 12, 2026

High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Coronary artery imaging with single-source rapid kilovolt peak-switching dual-energy CT
Jonathan A Scheske1, Julie M O'Brien, James P Earls
1Department of Medical Imaging, St. Paul's Hospital, 1081 Burrard St, Vancouver, BC, Canada, V6Z 1Y6.
Purpose:
To evaluate beam-hardening (BH) artifact reduction in coronary computed tomography (CT) angiography with dual-energy CT, to define the optimal monochromatic-energy levels for coronary and myocardial signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in dual-energy CT, and to compare these levels with single-energy CT.
Materials And Methods:
The study was approved by the institutional review board and/or ethics committee at each site. Patients provided informed consent. Thirty-nine patients were prospectively enrolled to undergo dual-energy CT, and 25 also underwent single-energy CT. Myocardial and coronary SNR, CNR, and iodine concentration were measured across multiple segments at varying monochromatic energy levels (40-140 keV). BH was defined as signal decrease in basal inferior wall versus midinferior wall, and signal increase in midseptum versus midinferior wall. Generalized estimating equation was used to identify optimal monochromatic-energy levels and compare them with single-energy CT.
Results:
BH was noted at single-energy CT with basal inferior wall mean reduction of 19.7 HU ± 29.2 (standard deviation) and midseptum increase of 46.3 HU ± 36.3. There was reduction in this artifact at 90 keV or greater (1.7 HU ± 18.4 in basal inferior wall and 20.1 HU ± 37.5 in midseptum at 90 keV; P < .05). SNR and CNR were higher in the myocardium and coronary arteries at 60-80 keV than single-energy CT (myocardium: SNR, 3.02 vs 2.39, and CNR, 6.73 vs 5.16; coronary arteries: SNR, 10.83 vs 7.75, and CNR, 13.31 vs 9.54; P < .01). Mean iodine concentration in resting myocardium was 2.19 mg/mL ± 0.57.
Conclusion:
Rapid kilovolt peak-switching dual-energy CT resulted in significant BH reduction and improvements in SNR and CNR in the myocardium and coronary arteries.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Acute Coronary Syndrome III: Diagnostic Studies
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Computed Tomography