Three-dimensional rotational angiography is preferable to conventional two-dimensional techniques for uterine artery
Atul Gupta1, Kirsten Zuurmond, Thijs Grünhagen
1Department of Interventional Radiology, Paoli Hospital, Paoli, PA, USA. guptaa@mlhs.org
Purpose:
We aimed to investigate the potential benefits of three-dimensional rotational angiography (3DRA) compared to two-dimensional (2D) roadmapping to visualize the uterine artery (UA) origins during uterine artery embolization (UAE) procedures.
Materials And Methods:
Sixty-three UAE cases performed under 3DRA guidance were reviewed retrospectively to determine if there was an optimal angiographic projection angle for identifying UA origin. Digital subtraction angiogram (DAS)-like images of the pelvic vessels were generated from the 3DRA scans at six different angles: left anterior oblique (LAO) 25°, 35°, 45°; and right anterior oblique (RAO) 25°, 35°, 45°. Two experienced interventional radiologists assessed if these angles could effectively serve as a roadmap to guide catheterization of the UA. Assessment was validated against original 3DRA scans to determine the percentage of true and false positives.
Results:
No single projection angle was found that could consistently be utilized for UA catheterization. The projection angles used during 3D roadmapping for both the left and right UA showed two clusters with both a wide spread: RAO 20° to 50° and LAO 20° to 50°. More than 50% of the DSA-like images at RAO and LAO 45° appeared to be adequate for UA catheterization, but validation against the 3DRA revealed 28% of these images were suboptimal and deceptive, due to unappreciated overlapping vessels.
Conclusion:
No standard projection angles can be recommended with 2D roadmapping to consistently visualize UA origin. The 3DRA can be as a useful tool for UAE to achieve reliable and consistent visibility of the UA origin.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies VII: Vascular Imaging
Venous Thrombosis III: Interprofessional Care