Can tibial coverage in total knee replacement be reliably evaluated with three-dimensional image-based digital
A J Costa1, S Lustig, C J Scholes
1Sydney Orthopaedic Research Institute, Chatswood, New South Wales, Australia.
Objectives:
There remains a lack of data on the reliability of methods to estimate tibial coverage achieved during total knee replacement. In order to address this gap, the intra- and interobserver reliability of a three-dimensional (3D) digital templating method was assessed with one symmetric and one asymmetric prosthesis design.
Methods:
A total of 120 template procedures were performed according to specific rotational and over-hang criteria by three observers at time zero and again two weeks later. Total and sub-region coverage were calculated and the reliability of the templating and measurement method was evaluated.
Results:
Excellent intra- and interobserver reliability was observed for total coverage, when minimal component overhang (intraclass correlation coefficient (ICC) = 0.87) or no component overhang (ICC = 0.92) was permitted, regardless of rotational restrictions.
Conclusions:
Measurement of tibial coverage can be reliable using the templating method described even if the rotational axis selected still has a minor influence.