A comparison of two ambulatory blood pressure monitors worn at the same time
Radhakrishna R Kallem1, Kevin E C Meyers, Deirdre L Sawinski
1Renal, Electrolyte and Hypertension Division, Department of Medicine, University of Pennsylvania School of Medicine, Philadelphia, PA 19104, USA. krishna.kallem@uphs.upenn.edu
Abstract:
There are limited data in the literature comparing two simultaneously worn ambulatory blood pressure (BP) monitoring (ABPM) devices. The authors compared BPs from two monitors (Mobil-O-Graph [I.E.M., Stolberg, Germany] and Spacelabs 90207 [Spacelabs Medical, Issequah, WA]). In the nonrandomized component of the study, simultaneous 8-hour BP and heart rate data were measured by Mobil-O-Graph, consistently applied to the nondominant arm, and Spacelabs to the dominant arm on 12 untreated adults. Simultaneous 8-hour BP and heart data were obtained by the same monitors randomly assigned to a dominant or nondominant arm on 12 other untreated adults. Oscillometric BP profiles were obtained in the dominant and nondominant arms of the above 24 patients using an Accutorr (Datascope, Mahwah, NJ) device. The Spacelabs monitor recorded a 10.2-mm Hg higher systolic pressure in the nonrandomized (P=.0016) and a 7.9-mm Hg higher systolic pressure in the randomized studies (P=.00008) compared with the Mobil-O-Graph. The mean arterial pressures were 1 mm Hg to 2 mm Hg different between monitors in the two studies, and heart rates were nearly identical. Our observations, if confirmed in larger cohorts, support the concern that ABPM device manufacturers consider developing normative databases for their devices.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: