An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Bone microstrain values of 1-piece and 2-piece implants subjected to mechanical loading
Noga Harel1, Inbar Eshkol-Yogev, Dana Piek
1Department of Oral Rehabilitation, The Maurice and Gabriela Goldschleger School of Dental Medicine, Tel Aviv University, Tel Aviv 69978, Israel. noga.hrl@gmail.com
Purpose:
The purpose of this study was to measure and compare the strain levels in peri-implant bone as generated by 1-piece (1P) and 2-piece (2P) implant systems.
Materials And Methods:
The implants (1P and 2P) were placed into bovine bone according to the manufacturer's protocol. Four linear strain gauges were placed around each implant neck and apex. Each model was loaded in static loading by a material testing machine in ascending forces ranging from 20 to 120 N. Microstrains (μ[Latin Small Letter Open E]) generated in the surrounding bone were measured by a strain gauge and recorded.
Results:
Recorded microstrains were significantly higher for 1P implants than for 2P implants. Average recorded microstrain values were significantly lower in the neck (71.6 and 17.3 µs) compared with the apical (132 and 60 µs) regions of 1P and 2P implants, respectively (P < 0.0001).
Conclusions:
Within the limitations of this study, highest microstrains were generated in apical regions regardless of implant design, but the 2P implant ap-peared to provide a stress-damping effect in both the cervical and apical regions compared with the 1P implant.
Related Concept Videos
Normal Strain under Axial Loading
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Plastic Behavior
True Stress and True Strain
In contrast, true stress offers a more precise portrayal. It is computed by dividing the...
Strain and Elastic Modulus
Stress-Strain Diagram - Ductile Materials