Related Experiment Videos
Electrochemical detection of the 3,5-dinitrobenzoyl derivative of digoxin by high-performance liquid chromatography
1Faculty of Pharmaceutical Sciences, University of British Columbia, Vancouver, Canada.
Journal of Chromatography
|April 6, 1990
Abstract:
Electrochemical detection of 3,5-dinitrobenzoyl derivatives of digoxin and its metabolites following high-performance liquid chromatography is reported. Partial resolution of derivatized digoxin and dihydrodigoxin was obtained using a Spherisorb ODS II analytical column. Both single- and dual-electrode detection were investigated and a maximum sensitivity equivalent to 0.39 ng of digoxin was found with the dual-electrode method. This system has the necessary sensitivity and selectivity for development into a therapeutic monitoring assay method.