Effect of an orthosis on the loads acting on a vertebral body replacement
Antonius Rohlmann1, Thomas Zander, Friedmar Graichen
1Julius Wolff Institute, Charité-Universitätsmedizin Berlin, Augustenburger Platz 1, 13353 Berlin, Germany. antonius.rohlmann@charite.de
Background:
The spinal load reduction by an orthosis is still a matter of debate. Some studies predicted a load reduction while others found no effect. The aim of this study was to measure the in vivo effect of the Lumbo TriStep brace and the hyperextension orthosis medi 3C on the spinal implant loads.
Methods:
Telemeterized vertebral body replacements were implanted in 5 patients suffering from a severe fracture of the L1 or L3 vertebral body. The implant allows the measurement of 6 load components acting on it. For several activities during standing, sitting and walking, implant loads were measured in patients with and without an orthosis.
Findings:
The average resultant force on the vertebral body for 26 activities was reduced by 9% with the Lumbo TriStep brace, and by 19% with the hyperextension orthosis. The force reduction is usually more pronounced for activities performed during sitting than it is for those performed while standing. However, considerable inter- and intra-individual variation was observed. In several cases, the measured implant forces were even higher when the patients were wearing an orthosis.
Interpretation:
In some patients, for certain activities, an orthosis may reduce the force on a vertebral body replacement and thus on the anterior column of the spine. However, in other patients for the same activities, an orthosis may increase the force. The measurements do not allow a clear recommendation to wear an orthosis since the clinically relevant reduction of implant forces is unknown.
Related Concept Videos
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical bending,...
Eccentric Loading
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Impact Loading
In cases of elastic deformation,...
General Structure of a Vertebra
Eccentric Axial Loading in a Plane of Symmetry