Related Experiment Video
Updated: May 11, 2026

Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
A new shallow underground gas-proportional counting lab--first results and Ar-37 sensitivity
C E Aalseth1, A R Day, E S Fuller
1Pacific Northwest National Laboratory, 902 Battelle Boulevard, P.O. Box 999, MSIN J4-65, Richland, WA 99352, USA. craig.aalseth@pnnl.gov
Abstract:
A new ultra-low-background proportional counter was recently developed with an internal volume of 100 cm(3) and has been characterized at pressures from 1-10 atm with P-10 (90% Ar, 10% methane) gas. This design, along with a counting system providing event digitization and passive and active shielding, has been developed to complement a new shallow underground laboratory (30 m water-equivalent). Backgrounds and low-level reference materials have been measured, and system sensitivity for (37)Ar has been calculated.