Effect of hydroxocobalamin on surface oximetry in nonexposed humans
Brian V Cashin1, Aaron G Matlock, Chris Kang
1Department of Emergency Medicine, Madigan Army Medical Center, Tacoma, WA, USA.
Introduction:
The newer cyanide antidote, hydroxocobalamin, due to its pigmentation, has been found to cause interferences in some laboratory assays. Co-oximetry may also be affected by hydroxocobalamin, leading to false elevations in hemoglobin concentration, methemoglobin, carboxyhemoglobin, and false decreases in oxyhemoglobin. The Masimo Radical-7 is a medical device that performs noninvasive oximetry and estimates hemoglobin (Hb) concentration and percent carboxyhemoglobin (COHb), methemoglobin (MetHb), and oxyhemoglobin saturation (O2Hb).
Study Objectives:
The study sought to determine the effect of hydroxocobalamin on noninvasive measurement of hemoglobin indices using the Masimo Radical-7 monitor.
Methods:
Seven asymptomatic volunteers who were unexposed to cyanide had baseline heart rate (HR), blood pressure (BP), and oximeter measurements recorded followed by an infusion of five grams of hydroxocobalamin over 15 minutes. The above parameters were subsequently recorded at: 5, 10, 15, 30 and 60 minutes post infusion. Data were analyzed by calculating the area under the curve (AUC) for each variable and comparing the results to expected values by paired t tests. Expected AUC values were calculated by extrapolating baseline values across the entire time period.
Results:
The mean differences from baseline values with 95% confidence intervals and t tests of mean difference were: SBP: 11 mm Hg (95% CI, 0-22; P = .051); HR: -9 (95% CI, -15 to -3; P = .01); Hb: -0.1 (95% CI, -0.7 to 0.4; P = .57); O2Hb: 0 g/dL (95% CI, -1 to 1; P = .41); COHb: -1 (95% CI, -3 to 1; P = .25); MetHb: -0.2 (95% CI, -0.3 to 0; P = .03).
Discussion:
After infusion of hydroxocobalamin there was a significant elevation of systolic blood pressure and decrease in heart rate. There were no significant differences in Hb, O2Hb, and COHb. Although percent methemoglobin concentrations were statistically lower, the authors feel this difference is of trivial clinical significance.
Conclusion:
The administration of hydroxocobalamin does not significantly impact noninvasive oximetry.
Related Concept Videos
Oxygen Transport in the Blood
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...