Treatment of buffy coat platelets in platelet additive solution with the mirasol(®) pathogen reduction technology

Azucena Castrillo1, Marcia Cardoso, Lindsay Rouse

  • 1Centro de Transfusion de Galicia, Santiago de Compostela, Spain.

Abstract

Insights

Mirasol pathogen reduction technology (PRT) effectively treats buffy coat platelet concentrates in platelet additive solution (PAS), maintaining acceptable cell quality for 7-day storage. Treated platelets show enhanced metabolism and activation within normal ranges.

Area of Science:

  • Blood banking and transfusion medicine
  • Biotechnology and medical devices
  • Hematology

Background:

  • Mirasol pathogen reduction technology (PRT) uses riboflavin and UV light to reduce pathogen load and inactivate leukocytes in blood products.
  • Currently approved in Europe for treating platelets and plasma.
  • Evaluates Mirasol system's impact on buffy coat platelet concentrates (BCPCs) in platelet additive solution (PAS).

Purpose of the Study:

  • To assess the quality of buffy coat platelet concentrates (BCPCs) prepared with PAS and treated with the Mirasol system.
  • To examine the effects of Mirasol treatment on platelet cell quality during a 7-day storage period.

Main Methods:

  • 26 BCPCs were prepared and divided into paired control and test units.
  • Test units underwent Mirasol system treatment.
  • Platelet quality was assessed over 7 days of storage.

Main Results:

  • All products met Mirasol treatment specifications.
  • Mirasol-treated BCPCs in PAS maintained required pH throughout storage.
  • Analysis indicated enhanced cellular metabolism and activation in treated platelets, with levels within established ranges.

Conclusions:

  • Mirasol-treated BCPCs in PAS exhibit increased metabolism and activation compared to controls.
  • These treated platelet units maintain acceptable cell quality for 7-day storage.
  • Mirasol PRT is a viable option for enhancing blood product safety.