Related Experiment Video
Updated: May 11, 2026

Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
Disc impediment in medtronic-Hall aortic valve prosthesis followed by successful reoperation
Sigita Glaveckaitė1, Ričardas Peldžius, Arimantas Grebelis
1Department of Cardiovascular Medicine, Faculty of Medicine, Vilnius University, Lithuania. sigita.glaveckaite@santa.lt
Abstract:
This case report describes the diagnostic and treatment challenges of the combined disc impediment caused by pannus ingrowth in combination with thrombus formation in a 21-mm Medtronic-Hall aortic valve prosthesis. A combined obstruction of a Medtronic-Hall prosthesis in the aortic position is a rare, but potentially fatal clinical condition for the first time reported in Vilnius University Hospital Santariškių Klinikos. Despite the difficulties in determining a correct diagnosis, our critically ill patient was clinically stabilized by partially successful thrombolysis and then successfully treated by a surgical thrombectomy together with pannus excision during the rotation of the disc within the valve housing 3 years after the valve implantation.