Related Experiment Video
Updated: May 11, 2026

Design of a Biocompatible Drug-Eluting Tracheal Stent in Mice with Laryngotracheal Stenosis
Published on: January 21, 2020
Biodegradable polymer biolimus-eluting stent versus durable polymer everolimus-eluting stent: a randomized,
Masahiro Natsuaki1, Ken Kozuma2, Takeshi Morimoto3
1Department of Cardiovascular Medicine, Kyoto University Graduate School of Medicine, Kyoto, Japan.
Objectives:
NEXT (NOBORI Biolimus-Eluting Versus XIENCE/PROMUS Everolimus-Eluting Stent Trial) was designed for evaluating the noninferiority of a biolimus-eluting stent (BES) relative to an everolimus-eluting stent (EES) in terms of target lesion revascularization (TLR) at 1 year.
Background:
Efficacy and safety data comparing biodegradable polymer BES with durable polymer cobalt-chromium EES are currently limited.
Methods:
The NEXT trial is a prospective, multicenter, randomized, open-label, noninferiority trial comparing BES with EES. Between May and October 2011, 3,235 patients were randomly assigned to receive either BES (n = 1,617) or EES (n = 1,618).
Results:
At 1 year, the primary efficacy endpoint of TLR occurred in 67 patients (4.2%) in the BES group, and in 66 patients (4.2%) in the EES group, demonstrating noninferiority of BES relative to EES (p for noninferiority <0.0001, and p for superiority = 0.93). Cumulative incidence of definite stent thrombosis was low and similar between the 2 groups (0.25% vs. 0.06%, p = 0.18). An angiographic substudy enrolling 528 patients (BES: n = 263, and EES: n = 265) demonstrated noninferiority of BES relative to EES regarding the primary angiographic endpoint of in-segment late loss (0.03 ± 0.39 mm vs. 0.06 ± 0.45 mm, p for noninferiority <0.0001, and p for superiority = 0.52) at 266 ± 43 days after stent implantation.
Conclusions:
One-year clinical and angiographic outcome after BES implantation was noninferior to and not different from that after EES implantation in a mostly stable coronary artery disease population. One-year clinical outcome after both BES and EES use was excellent, with a low rate of TLR and extremely low rate of stent thrombosis.
Related Concept Videos
Bioequivalence studies: Biowaivers
Site-Targeted Drug Delivery Systems: Polymeric Carriers
Bioavailability Enhancement: Drug Stability Enhancement and GI Retention
Bioequivalence Experimental Study Designs: Completely Randomized and Randomized Block Designs