Evaluation of a new nonnvasive device in determining hemoglobin levels in emergency department patients
Related Experiment Video
Updated: May 11, 2026

A Point-of-Care Method with Integrated Decision Support Tool to Estimate Anemia at Population Level
Published on: January 19, 2024
Tristan Knutson1, David Della-Giustina, Eric Tomich
1Madigan Army Medical Center, Department of Emergency Medicine, Tacoma, Washington.
Introduction:
The Masimo Radical-7 Pulse CO-Oximeter is a medical device recently approved by the US Food and Drug Administration that performs noninvasive oximetry and estimated venous or arterial hemoglobin measurements. A portable, noninvasive device that rapidly measures hemoglobin concentration could be useful in both austere and modern hospital settings. The objective of this study is to determine the degree of variation between the device's estimated hemoglobin measurement and the actual venous hemoglobin concentration in undifferentiated emergency department (ED) patients.
Methods:
We conducted a prospective, observational, cross-sectional study of adult patients presenting to the ED. The subjects consisted of a convenience sample of adult ED patients who required a complete blood count as part of their care in the ED. A simultaneous probe hemoglobin was obtained and recorded.
Results:
Bias between probe and laboratory hemoglobin measurements was -0.5 (95% confidence interval, - 0.8 to -0.1) but this was not statistically significant from 0 (t 0.05,124 = 0.20, P > 0.5). The limits of agreement were -4.7 and 3.8, beyond the clinically relevant standard of equivalency of ± 1 g/dL.
Conclusion:
These data suggest that noninvasive hemoglobin determination is not sufficiently accurate for emergency department use.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...