Single-shot speckle reduction and dispersion compensation in optical coherence tomography by compounding fractional
Norman Lippok1, Poul Nielsen, Frédérique Vanholsbeeck
1Physics Department, The University of Auckland, Private Bag, Auckland 92019, New Zealand. n.lippok@auckland.ac.nz
Abstract:
We present speckle suppression and dispersion compensation for Fourier-domain optical coherence tomography based on fractional Fourier transforms of a single A scan. A 1.54-fold reduction in speckle contrast was achieved with group velocity dispersion compensation. The method is demonstrated on biological samples using a swept source configuration at 1310 nm and a spectral-domain system at 840 nm.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Confocal Fluorescence Microscopy
Super-resolution Fluorescence Microscopy
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...