Cardiac three-dimensional rotational angiography can be performed with low radiation dose while preserving image
Stijn De Buck1, Becker S N Alzand, Jean-Yves Wielandts
1Department of Cardiovascular Sciences, Section of Arrhythmology, University Hospital Gasthuisberg, University of Leuven, Herestraat 49, B-3000 Leuven, Belgium, Europe.
Aims:
The effective radiation dose (ED) of three-dimensional rotational angiography (3DRA) is 5-8 mSv, leading to reticence on its use. We evaluated the potential of 3DRA with a reduced number of frames (RNF) and a reduced dose per frame.
Methods And Results:
Three-dimensional rotational angiography was performed in 60 patients (52.5 ± 9.6 years, 16 females) referred for ablation in the right (RA; n = 10) and left atrium (LA; n = 50). In a simulation group (n = 20), the effect of dropping frames from a conventional 248 frames 3DRA LA acquisition was simulated. In a prospective group (n = 40), RNF 3DRA were acquired of LA (n = 30) and RA (n = 10) with 67 frames (0.24 Gy/frame) and 45 frames (0.12 μGy/frame), respectively. Accuracy was evaluated qualitatively and quantitatively. Effective radiation dose was determined by Monte Carlo simulation on every frame. In the simulation group, surface errors increased minimally and non-significantly when reducing frames from 248 to 124, 83, 62, 50, 42, and 31: 0.49 ± 0.51, 0.52 ± 0.46, 0.61 ± 0.49, 0.62 ± 0.47, 0.71 ± 0.48, and 0.81 ± 0.47 mm, respectively (Pearson coefficient 0.20). All 3D LA images were clinically useful, even with only 31 frames. In the prospective group, good or optimal 3D image quality was achieved in 80% of LA and all of RA reconstructions. These accurate models were obtained with ED of 2.6 ± 0.4 mSv for LA and 1.2 ± 0.5 mSv for RA.
Conclusion:
Three-dimensional rotational angiography is possible with a significant reduction in ED (to the level of prospectively gated cardiac computed X-ray tomography) without compromising image quality. Low-dose 3DRA could become the preferred online 3D imaging modality for pulmonary vein isolation and other anatomy-dependent ablations.
More Related Videos
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System IV: CMRI