Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
Repeated thermo-sterilisation further affects the reliability of positive end-expiratory pressure valves
Julia Christine Hartung1, Georg Schmölzer, Gerd Schmalisch
1Department of Neonatology, Charité University Medical Center, Berlin, Germany.
Aim:
Positive end-expiratory pressure (PEEP) valves are used together with self-inflating bags (SIB) to provide a preset PEEP during manual ventilation. It has recently been shown that these valves deliver highly variable levels of PEEP. We hypothesised that material fatigue due to repeated thermo-sterilisation (TS) may contribute to varying reliability of PEEP valves.
Methods:
In a laboratory study 10 new PEEP valves were tested before and after 10, 20 and 30 cycles of routine TS (7 min at 134°C) by using a neonatal lung model (compliance 0.2 mL/kPa). Settings were positive inflation pressure = 20 and 40 cm H(2)O, PEEP = 5 and 10 cm H(2)O, respiratory rate = 40 and 60/min, flow = 8l/min. PEEP was recorded using a respiratory function monitor.
Results:
Before TS, a mean (standard deviation) PEEP of 4.0 (0.9) and 7.7 (1.0) cm H(2)O was delivered by the 10 valves when the PEEP was set to 5 and 10 cm H(2)O, respectively. One new valve only delivered 2.0 (0.0) and 5.0 (0.0) cm H(2)O when the PEEP was adjusted to 5 and 10 cm H(2)O, respectively. Four of the 10 investigated valves showed significant variations in PEEP (coefficient of variation >10%) throughout the autoclaving process. One valve completely lost its function after the 20th TS. Common defects were tears in the softer materials or displacement of the rubber seal. Six of the 10 valves continued to provide PEEP in spite of repeated TS.
Conclusion:
The reliability of PEEP valves is affected by repeated TS. Multi-use PEEP valves should be tested for reliable PEEP provision following TS.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Mechanical Ventilation I: Indication and Settings
Pneumothorax-II
Clinical Manifestations: