Innovative continuous non-invasive cuffless blood pressure monitoring based on photoplethysmography technology
Juan C Ruiz-Rodríguez1, Adolf Ruiz-Sanmartín, Vicent Ribas
1Critical Care Department, Vall d'Hebron University Hospital, Barcelona, Spain. jcruiz@vhebron.net
Purpose:
To develop and validate a continuous non-invasive blood pressure (BP) monitoring system using photoplethysmography (PPG) technology through pulse oximetry (PO).
Methods:
This prospective study was conducted at a critical care department and post-anesthesia care unit of a university teaching hospital. Inclusion criteria were critically ill adult patients undergoing invasive BP measurement with an arterial catheter and PO monitoring. Exclusion criteria were arrhythmia, imminent death condition, and disturbances in the arterial or the PPG curve morphology. Arterial BP and finger PO waves were recorded simultaneously for 30 min. Systolic arterial pressure (SAP), mean arterial pressure (MAP), and diastolic arterial pressure (DAP) were extracted from computer-assisted arterial pulse wave analysis. Inherent traits of both waves were used to construct a regression model with a Deep Belief Network-Restricted Boltzmann Machine (DBN-RBM) from a training cohort of patients and in order to infer BP values from the PO wave. Bland-Altman analysis was performed.
Results:
A total of 707 patients were enrolled, of whom 135 were excluded. Of the 572 studied, 525 were assigned to the training cohort (TC) and 47 to the validation cohort (VC). After data processing, 53,708 frames were obtained from the TC and 7,715 frames from the VC. The mean prediction biases were -2.98 ± 19.35, -3.38 ± 10.35, and -3.65 ± 8.69 mmHg for SAP, MAP, and DAP respectively.
Conclusions:
BP can be inferred from PPG using DBN-RBM modeling techniques. The results obtained with this technology are promising, but its intrinsic variability and its wide limits of agreement do not allow clinical application at this time.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.