Related Experiment Video
Updated: May 10, 2026

Transradial Access Chemoembolization for Hepatocellular Carcinoma Patients
Published on: September 20, 2020
Incidence of arterial micro-embolization during percutaneous AngioJet thrombectomy of hemodialysis grafts
Vasileios Koukounas1, Dimitris Karnabatidis, Athanasios Diamantopoulos
1Department of Interventional Radiology, School of Medicine, University of Patras, 26500, Rion, Greece.
Purpose:
This study was designed to investigate the incidence of arterial embolization using a peripheral protection filter device in a series of patients undergoing percutaneous mechanical thrombectomy for the management of thrombosed hemodialysis arteriovenous grafts (AVGs).
Methods:
This prospective, single-center study included all eligible patients presenting during an 18-month period to undergo AVG percutaneous thrombectomy. Inclusion criteria was a recently thrombosed AVG with 2 cm of artery before the next arterial branching. Primary endpoint was the incidence of distal arterial macro- and micro-embolization determined by both digital subtraction angiography and histopathological analysis of the material collected. Secondary endpoints included quantitative measurements of the specimens using a 0+ (no material) to 3+ (maximum load) score.
Results:
In total, 42 patients met the study's inclusion criteria. No procedure-related complications or angiographically evident arterial embolization were noted. Macroscopically evident material was present in 47.6% (20/42 filters). Histopathology demonstrated that the embolic material was primary consisted of fibrin conglomerates and platelets (median score: 1.5, confidence interval: 1.0-3.0), whereas inflammatory cells, trapped erythrocytes, extracellular matrix, cholesterol clefts, foam cells, necrotic core, and smooth muscle cells also were detected. Mean total area of embolic material was 5.04 mm(2) (range 0.05-5.21). The mean major axis of the largest particle was 1.83 mm (range 0.29-6.64), whereas 19% (8/42) contained particles with major axis >1 mm and 12% (5/42) with major axis >3 mm.
Conclusions:
In this study, the percentage of arterial micro-embolization was significantly higher than previously reported. However, the detrimental, long-term, clinical relevance remains to be determined.
Related Concept Videos
Hemodialysis I: Introduction
Hemodialysis II: Procedure and Complications
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care
Cardiac Catheterization I: Pre-Procedure Overview
Acute Kidney Injury IV: Diagnostic Studies and Prevention
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications