Related Experiment Video
Updated: May 10, 2026

Rapid Formation and Testing of Self-expanding NiTi Frames with a Small Form Factor Suitable for Minimally Invasive Implants
Published on: March 7, 2025
[The investigation of nickel dissolution from nitinol under different extraction conditions in biological tests]
Xiantao Wen1, Xueying He, Jian Li
1Beijing Institute of Medical Device Testing, Beijing, 101111. wenxiantao@hotmail.com
Abstract:
The dissolution of nickel from nitinol alloy under different extraction conditions was investigated when biological tests of nitinol medical devices were carried out. It discussed the equivalence of these extraction conditions and found the best extraction conditions. In the experiment, two brands of nitinol were chosen and extracted under different extraction conditions according to GB/T 16886-12. Plasma emission spectrometer was used to analyse the concentration of nickel in extract liquid. The results show that there is no equivalence among these extraction conditions for nitinol materials. And it is suggested that (37 +/- 1) degree C, (72 +/- 2) h could be used as normal extraction condition and (70 +/- 2) degrees C, (24 +/- 2) h as accelerated extraction condition.
Related Concept Videos
Precipitation Gravimetry
In determining nickel by gravimetric analysis, a precipitant of ethanolic dimethylglyoxime is added to a hot nickel salt solution. This is quickly followed by the dropwise addition of dilute ammonia solution until precipitation occurs. A...
Extraction: Advanced Methods