Monopolar cautery and adverse effects on cochlear implants.

Anita Jeyakumar1, Meghan Wilson, Jonathan E Sorrel

  • 1Department of Otorhinolaryngology, Louisiana State University, New Orleans, USA. ajeyak@lsuhsc.edu

JAMA Otolaryngology-- Head & Neck Surgery
|June 22, 2013
PubMed
Summary

Monopolar cautery did not damage cochlear implants or raise endocochlear temperature in a cadaveric study. This suggests monopolar cautery may be safe for patients with cochlear implants.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Recurrent Cochlear Implant Electrode Exposure in a Smoker With Prior Canal Wall Mastoidectomy: A Case Report.

Cureus·2026
Same author

"Make it a childhood memory and not a therapeutic memory": Young people's experience of early childhood intervention in Australia.

Research in developmental disabilities·2026
Same author

Cochlear Implants and Electrocautery: A Live Animal Model.

The Laryngoscope·2026
Same author

Impact of anesthetics choice on auditory brainstem response testing in children: A systematic review.

International journal of pediatric otorhinolaryngology·2026
Same author

Alternative Modalities for Temporal Bone Training in Otolaryngology: A Systematic Review.

Journal of otology·2025
Same author

Manual of Pediatric Balance Disorders, Second Edition: Robert C. O'Reilly, Terry Morley, Jacob R. Brodsky, Sharon L. Cushing; San Diego, CA: Plural Publishing.

Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology·2025

Area of Science:

  • Otolaryngology
  • Biomedical Engineering
  • Neurosurgery

Background:

  • Monopolar cautery is often contraindicated for cochlear implant patients due to presumed device risk.
  • Limited data exists to support this contraindication for cochlear implant use.

Purpose of the Study:

  • To assess the impact of monopolar cautery on cochlear implant device integrity.
  • To determine if monopolar cautery affects endocochlear temperature in implanted cochleae.

Main Methods:

  • Monopolar cautery was applied to cochlear implants in cadavers at varying settings (10W, 50W) for 30 minutes.
  • Impedance, integrity, and endocochlear temperature were monitored.
  • Post-explantation device failure analysis was conducted.

Related Experiment Videos

Last Updated: May 10, 2026

Performing Repeated Intraoperative Impedance Telemetry Measurements during Cochlear Implantation
06:54

Performing Repeated Intraoperative Impedance Telemetry Measurements during Cochlear Implantation

Published on: August 4, 2023

Main Results:

  • No changes in impedance or integrity were observed during or after cautery.
  • Endocochlear temperature remained stable, even at 50W settings.
  • Post-study device analysis revealed no evidence of damage.

Conclusions:

  • Monopolar cautery, at tested settings and durations, did not cause detectable damage to cochlear implants.
  • No significant endocochlear temperature changes were recorded.
  • Findings challenge the routine contraindication of monopolar cautery in cochlear implant patients.