Paper-based chemiresistor for detection of ultralow concentrations of protein
Marta Pozuelo1, Pascal Blondeau, Marta Novell
1Department of Analytical and Organic Chemistry, Universitat Rovira i Virgili, Campus Sescelades, C/Marcel·lí Domingo s/n, 43007 Tarragona, Catalonia, Spain.
Abstract:
A new paper-based chemiresistor composed of a network of single-wall carbon nanotubes (SWCNTs) and anti-human immunoglobulin G (anti-HIgG) is reported herein. SWCNTs act as outstanding transducers because they provide high sensitivity in terms of resistance changes due to immunoreaction. As a result, the resistance-based biosensor reaches concentration detection as low as picomolar. The resulting paper-based biosensor is sensitive, selective and employs low-cost substrate and simple manufacturing stages. Since chemiresistors require low-power equipment and are able to detect low concentrations with inexpensive materials, the present approach may pave the way for the development of resistive biosensors at very low-cost with high performances.