Defining hazards of supplemental oxygen therapy in neonatology using the FMEA tool
Anne Catherine van der Eijk1, Denise Rook, Jenny Dankelman
1Department of BioMechanical Engineering, Faculty of Mechanical, Maritime & Materials Engineering, Delft University of Technology, Mekelweg 2, 2628CD, Delft, The Netherlands. a.c.vandereijk@tudelft.nl
Purpose:
To prospectively evaluate hazards in the process of supplemental oxygen therapy in very preterm infants hospitalized in a Dutch NICU.
Methods:
A Failure Mode and Effects Analysis (FMEA) was conducted by a multidisciplinary team. This team identified, evaluated, and prioritized hazards of supplemental oxygen therapy in preterm infants. After accrediting "hazard scores" for each step in this process, recommendations were formulated for the main hazards.
Results:
Performing the FMEA took seven meetings of 2 hours. The top 10 hazards could all be categorized into three main topics: incorrect adjustment of the fraction of inspired oxygen (FiO2), incorrect alarm limits for SpO2, and incorrect pulse-oximetry alarm limits on patient monitors for temporary use. The FMEA culminated in recommendations in both educational and technical directions. These included suggestions for (changes in) protocols on alarm limits and manual FiO2 adjustments, education of NICU staff on hazards of supplemental oxygen, and technical improvements in respiratory devices and patient monitors.
Conclusions:
The FMEA prioritized flaws in the process of supplemental oxygen therapy in very preterm infants. Thanks to the structured approach of the analysis by a multidisciplinary team, several recommendations were made. These recommendations are currently implemented in the study's center.
Related Concept Videos
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...