Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Calculation method using Clarkson integration for the physical dose at the center of the spread-out Bragg peak in
Minoru Tajiri1, Takamasa Maeda, Yusuke Koba
1Department of Hospital, Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, 4-9-1 Anagawa, Inage-ku, Chiba 263-8555, Japan. m_tajiri@nirs.go.jp
Purpose:
In broad-beam carbon-ion radiotherapy performed using the heavy-ion medical accelerator in Chiba, the number of monitor units is determined by measuring the physical dose at the center of the spread-out Bragg peak (SOBP) for the treatment beam. The total measurement time increases as the number of treatment beams increases, which hinders the treatment of an increased number of patients. Hence, Kusano et al. [Jpn. J. Med. Phys. 23(Suppl. 2), 65-68 (2003)] proposed a method to calculate the physical dose at the center of the SOBP for a treatment beam. Based on a recent study, the authors here propose a more accurate calculation method.
Methods:
The authors measured the physical dose at the center of the SOBP while varying the circular field size and range-shifter thickness. The authors obtained the physical dose at the center of the SOBP for an irregularly shaped beam using Clarkson integration based on these measurements.
Results:
The difference between the calculated and measured physical doses at the center of the SOBP varied with a change in the central angle of the sector segment. The differences between the calculated and measured physical doses at the center of the SOBP were within ± 1% for all irregularly shaped beams that were used to validate the calculation method.
Conclusions:
The accuracy of the proposed method depends on both the number of angular intervals used for Clarkson integration and the fineness of the basic data used for calculations: sampling numbers for the field size and thickness of the range shifter. If those parameters are properly chosen, the authors can obtain a calculated monitor unit number with high accuracy sufficient for clinical applications.
Related Concept Videos
Biological Effects of Radiation
Calculation of Electric Flux
Radiation Pressure: Problem Solving
The average value of the rate of momentum transfer divided by the absorbing area represents the average force per...
Dose Size and Dosing Frequency: Determination Methods
Absorption of Radiation
Gravitational Potential Energy for Extended Objects