Comparison of SPME and static headspace analysis of blood alcohol concentration utilizing two novel chromatographic stationary phases
Related Experiment Video
Updated: May 9, 2026

Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
Jessica L Westland1, Frank L Dorman
1The Pennsylvania State University, 107 Whitmore Laboratories, University Park, PA 16802, USA. jlw1120@psu.edu
Abstract:
Headspace gas chromatography, coupled to flame ionization detection (GC-FID) analysis of blood alcohol concentration is a routine analysis carried out in forensic laboratories. A common concern with this analysis is the possible co-elution of a variety of other commonly encountered compounds with the target compound ethanol. By use of application specific columns, improved separation of ethanol, as well as the other potential components can be achieved. The presented method includes the evaluation of blood alcohol concentration by both direct gas headspace and SPME utilizing a new combination of GC columns. An investigation of method detection limits (MDLs) was also conducted in order to determine a reporting limit as well as the degree of uncertainty at the common threshold value of 0.08 g/dL. The study showed that under the conditions of this work, static headspace analysis with the use of t-butanol as an internal standard provided the most accurate and precise data with an MDL of 0.002 g/dL for the Rtx(®) - BAC PLUS 1 column and 0.005 g/dL for the Rtx(®) - BAC PLUS 2 column. The study also showed that the SPME analysis using a Carboxen/PDMS (65 μm) fiber with the use of t-butanol provided the lowest overall MDL of 0.0006 g/dL for both Rtx(®) - BAC PLUS 1 and 2 columns without loss of accuracy or precision.
Related Concept Videos
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Chromatography: Introduction
The phase in which the compounds linger or on which the compounds adsorb is called the stationary phase, whereas the mobile phase is the solvent that carries the solutes to be analyzed. In traditional column chromatography, the mixture flows through the stationary phase, and the compounds partition between the stationary and mobile phases...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Chromatographic Methods: Classification
Chromatographic techniques are typically named by...