Prevention of airway fires: do not overlook the expired oxygen concentration
Matthew Remz1, Isaac Luria, Michael Gravenstein
1From the Department of Anesthesiology, University of Florida College of Medicine, Gainesville, Florida.
Background:
It is generally accepted that when an ignition source is used the inspired oxygen concentration (FIO2) should be <30% in the breathing circuit to help prevent airway fires. The time and conditions required to reduce a high O2% in the breathing circuit to <30% has not yet been systematically studied.
Methods:
We evaluated the inspired and expired circuit oxygen concentration response times of an Aestiva Avance S/5 anesthesia machine to reach an FIO2 of <30% from a starting FIO2 of 100% and 60% after reducing the FIO2 to 21%. The circuit was connected to a human patient simulator which has a functional residual capacity of 2 L, total lung capacity of 2.8 L, an oxygen consumption of 200 mL/min, and respiratory quotient of 0.8. Fresh gas flow (FGF) inputs of 2 L/min and 5 L/min were chosen to represent a spectrum of typical clinical FGF rates. Minute ventilation was set at 4 L/min. Determining the requisite median time to reach an O2 concentration of <30% in the breathing circuit was the primary aim of the study.
Results:
The median times (1st-99th percent confidence interval) required to achieve inspiratory and expiratory oxygen concentrations of <30% with the extended circuit configuration when starting at 60% for 5 L FGFs were 35 (32-36) and 104 (88-122) seconds, respectively. With 2 L FGF, these median times increased to 303 (291-313) and 255 (232-278) seconds, respectively. A shortened circuit configuration (P = 0.006) and higher FGF flow rate (P < 0.0001) were noted to be factors decreasing the median time required to achieve an oxygen concentration of <30%.
Conclusions:
Both inspired and expired circuit oxygen concentration may take minutes to decrease to <30% depending on circuit length, FGF rate, and starting circuit oxygen concentration. During the reduction in FIO2, the expiratory oxygen concentration may be >30% for a considerable time after the FIO2 is in a "safe" range. An increased expired oxygen concentration should also be considered an airway fire risk, and patient care protocols may need to be modified based on future studies.
More Related Videos
Related Concept Videos
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Cardiopulmonary Resuscitation II: ACLS Airway Management
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.