Background:
Despite positive efficacy, thermotherapy is not widely used in clinical oncology. Difficulties associated with field penetration and controlling power deposition patterns in heterogeneous tissue have limited its use for heating deep in the body. Heat generation using iron-oxide super-paramagnetic nanoparticles excited with magnetic fields has been demonstrated to overcome some of these limitations. The objective of this preclinical study is to investigate the feasibility of treating bladder cancer with magnetic fluid hyperthermia (MFH) by analyzing the thermal dosimetry of nanoparticle heating in a rat bladder model.
Methods:
The bladders of 25 female rats were injected with 0.4 ml of Actium Biosystems magnetite-based nanoparticles (Actium Biosystems, Boulder CO) via catheters inserted in the urethra. To assess the distribution of nanoparticles in the rat after injection we used the 7 T small animal MRI system (Bruker ClinScan, Bruker BioSpin MRI GmbH, Ettlingen, Germany). Heat treatments were performed with a small animal magnetic field applicator (Actium Biosystems, Boulder CO) with a goal of raising bladder temperature to 42°C in <10min and maintaining for 60min. Temperatures were measured throughout the rat with seven fiberoptic temperature probes (OpSens Technologies, Quebec Canada) to characterize our ability to localize heat within the bladder target.
Results:
The MRI study confirms the effectiveness of the catheterization procedure to homogenously distribute nanoparticles throughout the bladder. Thermal dosimetry data demonstrate our ability to controllably raise temperature of rat bladder ≥1°C/min to a steady-state of 42°C.
Conclusion:
Our data demonstrate that a MFH system provides well-localized heating of rat bladder with effective control of temperature in the bladder and minimal heating of surrounding tissues.
More Related Videos
09:01Magnetic-, Acoustic-, and Optical-Triple-Responsive Microbubbles for Magnetic Hyperthermia and Pothotothermal Combination Cancer Therapy
Published on: May 22, 2020
13:41Magnetic Resonance-Guided High Intensity Focused Ultrasound Generated Hyperthermia: A Feasible Treatment Method in a Murine Rhabdomyosarcoma Model
Published on: January 13, 2023
Related Concept Videos
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...