Related Experiment Video
Updated: May 9, 2026

Application of Design Aspects in Uniaxial Loading Machine Development
Published on: September 19, 2018
The effects of loading on the preload and dimensions of the abutment screw for a 3-unit cantilever-fixed prosthesis
Gaurav Setia1, Hoda Yousef, David Ehrenberg
1Restorative Dentistry, University of California at San Francisco, San Francisco, CA, USA.
Objective:
The purpose of this study was to use an in vitro model system to compare the effects on the screw torque and screw dimensions within 2 commercially available implant systems from occlusal loading on a cantilevered-fixed partial denture.
Materials And Methods:
Cantilevered implant-supported 3-unit prostheses with 2 premolar abutments and 1 premolar pontic (7.3 mm in length) were made on resin casts containing 2 implant analogs for 2 implant systems: BioLok Silhouette Tapered Implant System (Birmingham, AL) and Zimmer Tapered Screw-Vent Implant System (Carlsbad, CA) with 10 samples in each group. Each sample was loaded with either of 2 protocols: (1) a load of 50 N on the cantilevered pontic unit and (2) a loading of 150 N on all 3 units. The outcome measures were (1) changes in residual torque of the abutment screws and (2) changes in screw dimension.
Results:
The BioLok Silhouette Tapered Implant group demonstrated slight but statistically significant torque loss 18.8% to 28.5% in both abutment screws for both protocols, P ≤ 0.05, without any changes in screw dimension. In the Zimmer Tapered Screw-Vent Implant group, there was a significant elongation of the abutment screws and a markedly significant 44.4%, (P ≤ 0.01) loss in torque in the mesial screw and a 28.5%, (P ≤ 0.05) loss in torque in the distal screw when the cantilever alone was loaded.
Conclusions:
Differences in screw design influence the maintenance of preload and distortion of the shank. The influence of the interface design, namely an internal hex of 1 mm versus an external hex did not influence the preload. Cantilevered prostheses can cause loss of torque and dimensional changes in abutment screws.
Related Concept Videos
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Applications of Stress
The...
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Eccentric Loading
Residual Stresses in Bending