Related Experiment Video
Updated: May 9, 2026

Thermal Ablation for the Treatment of Abdominal Tumors
Published on: March 7, 2011
Direct thermography-a new in vitro method to characterize temperature kinetics of ablation catheters
M Fiek1, F Gindele, C von Bary
1Medical Hospital I, Klinikum Pasing, Steinerweg 5, 81241, Munich, Germany, Michael.Fiek@med.uni-muenchen.de.
Purpose:
For the treatment of increasingly complex cardiac arrhythmias, new catheter designs as well as alternative energy sources are constantly being developed. However, there is presently no in vitro method available for assessment of the temperature changes induced at various myocardial levels during energy delivery. Therefore, our study was aimed at developing an in vitro model to record and display the temperature kinetics during ablation in the entire muscle cross section.
Methods And Results:
A sapphire glass pane was inserted into one wall of the in vitro experimental set-up. Due to its thermodynamic properties, the temperature distribution in an adjacent cross section of the cardiac muscle can be measured exactly ( 1 °C) through this pane by means of a thermography camera. Computer-supported image processing enables the colour-coded and two-dimensional display of the temperature kinetics during the energy application at any location of the myocardial cross section (± 0.5 mm). This new measuring methodology was validated by direct temperature measurements utilizing several intramyocardial thermo elements.
Conclusion:
This new method allows a temporal and spatial analysis of the temperature phenomena during ablation without the interference and spatial limitation of intramyocardial temperature probes. New ablation technologies can thus be evaluated, independent of the catheter configuration or source of energy used.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...