A low-power wide-range I-v converter for amperometric sensing applications
IEEE Transactions on Biomedical Circuits and Systems
|July 16, 2013
Summary
This study introduces a novel current-to-voltage converter for amperometric sensing, achieving low power consumption (9.82 µW) for wearable and implantable devices. Its design enables sensitive current measurement from 1 pA to 1 µA.
Area of Science:
- Electronics
- Sensor Technology
- Biomedical Engineering
Background:
- Amperometric sensing requires precise current measurement.
- Existing sensors often face limitations in power consumption, hindering miniaturization for wearable or implantable applications.
- Low-power electronics are crucial for the development of next-generation biomedical devices.
Purpose of the Study:
- To present the first experimental results of a novel current-to-voltage converter.
- To demonstrate its suitability for amperometric sensing across a wide current range (1 pA to 1 µA).
- To achieve ultra-low power consumption for integration into wearable or implantable sensors.
Main Methods:
- The circuit combines transistors operating in weak inversion with floating-gate metal-oxide semiconductor (FG-MOS) devices.
- Three distinct gain settings were implemented to optimize performance.
- The design strategy focused on minimizing power consumption.
Main Results:
- Experimental validation of the current-to-voltage converter was achieved.
- The converter successfully measured currents in the range of 1 pA to 1 µA.
- The circuit demonstrated ultra-low power consumption, measuring 9.82 µW under normal operation.
Conclusions:
- The developed current-to-voltage converter is a significant advancement for low-power amperometric sensing.
- Its design facilitates integration into power-constrained wearable and implantable sensor systems.
- This technology paves the way for more sophisticated and accessible biomedical monitoring devices.
Related Concept Videos
Amperometry: Overview
Amperometry is a technique commonly used to measure the concentration of specific analytes in a solution by monitoring the electric current generated during an electrochemical reaction. It involves applying a constant potential between a working electrode and a reference electrode to measure the resulting current, which is proportional to the concentration of the analyte. The Clark oxygen electrode operates based on this principle of amperometry. It consists of a cathode and an anode enclosed...
Potentiometry: Membrane Electrodes
Membrane electrodes, also known as p-ion electrodes, use membranes that selectively interact with free analyte ions, generating a potential difference across the membrane. The resulting membrane potential, known as the asymmetry potential, is not zero even when analyte concentrations on both sides of the membrane are equal. The membrane's response is typically not selective to a single analyte but proportional to the concentration of all ions in the sample solution capable of interacting at the...
Voltammetric Techniques: Linear-Scan (E vs Time)
Polarography is a classical voltammetric technique used to analyze electrochemical reactions. This method applies a linear potential sweep to a dropping mercury electrode (DME), and the resulting current is measured. A dropping mercury electrode is commonly used as the working electrode in polarography. It consists of a capillary tube filled with mercury, where the tiny droplet forms at the tip. This droplet continuously drops from the capillary, creating a new electrode surface for each...
Potentiometry: Overview
Potentiometry is an analytical technique that measures the potential difference between two electrodes in an electrochemical cell without drawing any significant current that could alter the solution's composition. This method employs an indicator electrode, which exchanges electrons with the analyte solution, and a reference electrode with a constant potential. Each electrode is immersed in a solution comprised of two half-cells. In a conventional setup, the reference electrode serves as the...
Potentiometry: Types of Electrodes
Reference electrodes serve as a stable reference point for potentiometric measurements, while indicator and working electrodes react to variations in the composition of a solution.
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Voltammetry: Overview
Voltammetry is an electroanalytical technique in which the current flowing through an electrochemical cell is measured as a function of applied potential, typically under conditions of concentration polarization. The technique provides valuable information about redox-active species, and the current response is plotted as a voltammogram.
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...