Patch testers' opinions regarding diagnostic criteria for metal hypersensitivity reactions to metallic implants
Peter C Schalock1, Jacob P Thyssen
1Department of Dermatology, Massachusetts General Hospital, Harvard School of Medicine, Boston, MA 02114, USA. schalock.prof@gmail.com
Background:
Metal hypersensitivity reactions to implanted devices remain a challenging and controversial topic. Diagnostic criteria and methods are not well delineated.
Objective:
Diagnostic criteria for hypersensitivity reactions after metallic device implantation are evaluated in this study by a multinational group of patch testers using Thyssen's previously published criteria.
Methods:
A total of 119 dermatologists at the 2012 European Contact Dermatitis Society and 2013 American Contact Dermatitis Society meetings answered a survey regarding their opinions on topics relating to metal hypersensitivity.
Conclusions:
Four major and 5 minor diagnostic criteria emerged. Approximately 80% of respondents found the following criteria useful (major criteria): chronic dermatitis beginning weeks to months after metallic implantation, eruption overlying the metal implant, positive patch test to a metal component of the implant, and complete clearing after removal of the potentially allergenic implant. Minor criteria (<61% of respondents) were as follows: systemic allergic dermatitis reaction, therapy-resistant dermatitis, morphology consistent with dermatitis, histology consistent with allergic contact dermatitis, and a positive in vitro test to metals (eg, lymphocyte transformation test). In the challenging situation such as a symptomatic or failing orthopedic device, applying these 4 major criteria and the 5 supportive minor criteria may be useful for guiding decision making.
More Related Videos
12:19Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo
Published on: July 1, 2013
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Hypersensitivity Reactions: Immune-Complex Reactions
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Pulmonary Tuberculosis IV
Several diagnostic approaches are used to detect TB. The conventional method is the Tuberculin Skin Test (TST), also known as the Mantoux test. However, this method has...