Intravascular enhancement with identical iodine delivery rate using different iodine contrast media in a circulation
Casper Mihl1, Joachim E Wildberger, Tomas Jurencak
1From the *Department of Radiology, †CARIM School for Cardiovascular Diseases, Maastricht University Medical Center, Maastricht, The Netherlands; ‡Health Informatics Research and Strategy, MEDRAD Radiology & Interventional, Bayer HealthCare, Indianola, PA; §Faculty of Health, Medicine and Life Sciences, Maastricht University, Maastricht, The Netherlands; Departments of ∥Neuroradiology, and ¶Nuclear Medicine, University Hospital RWTH Aachen, Aachen, Germany.
Purpose:
Both iodine delivery rate (IDR) and iodine concentration are decisive factors for vascular enhancement in computed tomographic angiography. It is unclear, however, whether the use of high-iodine concentration contrast media is beneficial to lower iodine concentrations when IDR is kept identical. This study evaluates the effect of using different iodine concentrations on intravascular attenuation in a circulation phantom while maintaining a constant IDR.
Materials And Methods:
A circulation phantom with a low-pressure venous compartment and a high-pressure arterial compartment simulating physiological circulation parameters was used (heart rate, 60 beats per minute; stroke volume, 60 mL; blood pressure, 120/80 mm Hg). Maintaining a constant IDR (2.0 g/s) and a constant total iodine load (20 g), prewarmed (37°C) contrast media with differing iodine concentrations (240-400 mg/mL) were injected into the phantom using a double-headed power injector. Serial computed tomographic scans at the level of the ascending aorta (AA), the descending aorta (DA), and the left main coronary artery (LM) were obtained. Total amount of contrast volume (milliliters), iodine delivery (grams of iodine), peak flow rate (milliliter per second), and intravascular pressure (pounds per square inch) were monitored using a dedicated data acquisition program. Attenuation values in the AA, the DA, and the LM were constantly measured (Hounsfield unit [HU]). In addition, time-enhancement curves, aortic peak enhancement, and time to peak were determined.
Results:
All contrast injection protocols resulted in similar attenuation values: the AA (516 [11] to 531 [37] HU), the DA (514 [17] to 531 [32] HU), and the LM (490 [10] to 507 [17] HU). No significant differences were found between the AA, the DA, and the LM for either peak enhancement (all P > 0.05) or mean time to peak (AA, 19.4 [0.58] to 20.1 [1.05] seconds; DA, 21.1 [1.0] to 21.4 [1.15] seconds; LM, 19.8 [0.58] to 20.1 [1.05] seconds).
Conclusions:
This phantom study demonstrates that constant injection parameters (IDR, overall iodine load) lead to robust enhancement patterns, regardless of the contrast material used. Higher iodine concentration itself does not lead to higher attenuation levels. These results may stimulate a shift in paradigm toward clinical usage of contrast media with lower iodine concentrations (eg, 240 mg iodine/mL) in individual tailored contrast protocols. The use of low-iodine concentration contrast media is desirable because of the lower viscosity and the resulting lower injection pressure.
Related Concept Videos
Imaging Studies III: Computed Tomography
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies VII: Vascular Imaging
Compartment Models: Two-Compartment Model
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions