Locked plating of comminuted distal femur fractures: does unlocked screw placement affect stability and failure?
Shari Cui1, J G Bledsoe, Heidi Israel
1Department of Orthopaedic Surgery, St Louis University, St Louis, MO.
Objectives:
Locked plates provide greater stiffness, possibly at the expense of fracture healing. The purpose of this study is to evaluate construct stiffness of distal femur plates as a function of unlocked screw position in cadaveric distal femur fractures.
Methods:
Osteoporotic cadaveric femurs were used. Four diaphyseal bridge plate constructs were created using 13-hole distal femur locking plates, all with identical condylar fixation. Constructs included all locked (AL), all unlocked (AUL), proximal unlocked (PUL), and distally unlocked (DUL) groups. Constructs underwent cyclic axial loading with increasing force per interval. Data were gathered on axial stiffness, torsional stiffness, maximum torque required for 5-degree external rotation, and axial force to failure.
Results:
Twenty-one specimens were divided into AL, AUL, PUL, and DUL groups. Axial stiffness was not significantly different between the constructs. AL and PUL demonstrated greater torsional stiffness, maximum torque, and force to failure than AUL and AL showed greater final torsional stiffness and failure force than DUL (P < 0.05). AL and PUL had similar axial, torsion, and failure measures, as did AUL and DUL constructs. All but 2 specimens fractured before medial gap closure during failure tests. Drop-offs on load-displacement curves confirmed all failures.
Conclusions:
Only the screw nearest the gap had significant effect on torsional and failure stiffness but not axial stiffness. Construct mechanics depended on the type of screw placed in this position. This screw nearest the fracture dictates working length stiffness when the working length itself is constant and in turn determines overall construct stiffness in osteoporotic bone.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
05:34A Mini-Invasive Internal Fixation Technique for Studying Immobilization-Induced Knee Flexion Contracture in Rats
Published on: May 20, 2019
Related Concept Videos
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...