Related Experiment Video
Updated: May 9, 2026

Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Computed tomography x-ray tube life analysis: a multiyear study
1Diagnostic X-Ray Quality Assurance Section, Department of Medical Physics, at Memorial Sloan-Kettering Cancer Center, New York City, USA.
Purpose:
To assess and compare actual computed tomography (CT) x-ray tube life with manufacturer warranty coverage limits because prolonging the tube life helps to lower operating costs.
Methods:
Ten GE LightSpeed CT scanners with 40 Performix Ultra tube changes and 3 GE VCT scanners with 10 Performix Pro tube changes were followed for 6 years. CT x-ray tube life measurements were performed by analyzing log files of the units after a tube change.
Results:
The Ultra tubes warranty coverage limit is 70 kAs or 12 months, whichever comes first. For Pro tubes, it is 6000 examinations or 12 months, whichever comes first. Measurements for the Performix Ultra CT x-ray tubes showed a range of 7 to 48 months and 16.7 to 239.9 kAs. Mean values for the Ultra CT x-ray tubes were 19.2 ± 12.5 months and 81.0 ± 45.4 kAs. Seven Ultra CT x-ray tubes did not meet the warranty coverage limits, with an average life of 8 months and 48.1 kAs. For the Pro CT x-ray tubes, the measured logs indicated 22.4 ± 9.6 months of CT x-ray tube life. All 10 Pro CT x-ray tubes exceeded company warranty coverage limits.
Discussion:
Although Pro tubes lasted longer, they acquired fewer scans than did Ultra tubes. A similar result was shown for current output.
Conclusion:
Because the clinical demand for a CT scanner varies, it is difficult to determine the reason for the failed tubes. Mechanical, environmental, and usage factors can reduce the life expectancy of an x-ray tube.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...