Implantable electrolyte conductance-based pressure sensing catheter, II. Device construction and testing
Robert Tan1, Peyman Benharash, Peter Schulam
1Department of Bioengineering, University of California Los Angeles, Los Angeles, CA, 90095, USA.
Abstract:
Direct measurements of arterial blood pressure most commonly use bulky external instrumentation containing a pressure transducer connected to an ex vivo fluid-filled arterial line, which is subject to several sensing artifacts. In situ blood pressure sensors, typically solid state piezoresistive, capacitive, and interferometric sensors, are unaffected by these artifacts, but can be expensive to produce and miniaturize. We have developed an alternative approach to blood pressure measurement based on deformation of an elastic tube filled with electrolyte solution. Simple measurement of the electrical conductance of this solution as the tube dimensions change allows determination of the external pressure. The sensor is made from inexpensive materials and its miniaturization is straightforward. In vitro static testing of initial sensor prototypes mounted on a catheter tip showed a linear response with applied pressure and a resolution of 1 mmHg. In vivo sensing followed catheterization of the sensor into the femoral artery of a porcine model through a 7F catheter port. The sensor performed comparably to a commercial pressure transducer also connected to the catheter port. Due to its scalability and cost, this sensor has the potential for use in a range of pressure sensing applications, such as measurement of intracranial, spinal, or interstitial pressures.