Related Experiment Video
Updated: May 9, 2026

High-throughput, Automated Extraction of DNA and RNA from Clinical Samples using TruTip Technology on Common Liquid Handling Robots
Published on: June 11, 2013
Automated, simple, and efficient influenza RNA extraction from clinical respiratory swabs using TruTip and epMotion
Sara B Griesemer1, Rebecca Holmberg, Christopher G Cooney
1Laboratory of Viral Diseases, Wadsworth Center, New York State Department of Health, 120 New Scotland Avenue, Albany, NY 12208, United States. sbg03@health.state.ny.us
Background:
Rapid, simple and efficient influenza RNA purification from clinical samples is essential for sensitive molecular detection of influenza infection. Automation of the TruTip extraction method can increase sample throughput while maintaining performance.
Objectives:
To automate TruTip influenza RNA extraction using an Eppendorf epMotion robotic liquid handler, and to compare its performance to the bioMerieux easyMAG and Qiagen QIAcube instruments.
Study Design:
Extraction efficacy and reproducibility of the automated TruTip/epMotion protocol was assessed from influenza-negative respiratory samples spiked with influenza A and B viruses. Clinical extraction performance from 170 influenza A and B-positive respiratory swabs was also evaluated and compared using influenza A and B real-time RT-PCR assays.
Results:
TruTip/epMotion extraction efficacy was 100% in influenza virus-spiked samples with at least 745 influenza A and 370 influenza B input gene copies per extraction, and exhibited high reproducibility over four log10 concentrations of virus (<1% CV). RNA yields between the three automated methods differed by less than 0.5 log10 gene copies. 99% of clinical specimens that were PCR-positive after easyMAG or QIAcube extraction were also positive following TruTip extraction. Overall Ct value differences obtained between TruTip/epMotion and easyMAG/QIAcube clinical extracts ranged from 1.24 to 1.91. Pairwise comparisons of Ct values showed a high correlation of the TruTip/epMotion protocol to the other methods (R2>0.90).
Conclusion:
The automated TruTip/epMotion protocol is a simple and rapid extraction method that reproducibly purifies influenza RNA from respiratory swabs, with comparable efficacy and efficiency to both the easyMAG and QIAcube instruments.