Ultrasound Assessment of Endothelial Function: A Technical Guideline of the Flow-mediated Dilation Test
Published on: April 27, 2016
A novel noninvasive and simple method for assessment of endothelial function: enclosed zone flow-mediated
Naomi Idei1, Teiji Ukawa, Masato Kajikawa
1Department of Cardiovascular Medicine, Hiroshima University Graduate School of Biomedical Sciences, Hiroshima, Japan.
Background:
It is clinically important to estimate the degree of endothelial dysfunction. Several methods have been used to assess endothelial function in humans. Recently, we developed a new noninvasive method for measurement of vascular response to reactive hyperemia in the brachial artery, named enclosed zone flow-mediated vasodilation (ezFMD). The purpose of this study was to determine the validity of ezFMD for assessment of endothelial function.
Methods And Results:
We measured ezFMD by a new device using an oscillometric method and conventional FMD using ultrasonography in 306 subjects, including patients with hypertension, dyslipidemia, and diabetes mellitus (218 men and 88 women, 30 ± 16 yr). Univariate regression analysis revealed that ezFMD significantly correlated with age (r = -0.42, P < 0.0001), body mass index (r = -0.13, P = 0.028), systolic blood pressure (r = -0.15, P = 0.009), diastolic blood pressure (r = -0.14, P = 0.011), fasting glucose level (r = -0.27, P = 0.006), smoking (r = -0.21, P = 0.007) and baseline pulse wave amplitude (r = -0.51, P < 0.0001). ezFMD significantly correlated with conventional FMD (r = 0.34, P < 0.0001). Multiple regression analysis revealed that age (P = 0.002), body mass index (P = 0.013), systolic blood pressure (P = 0.009), smoking (P = 0.004) and baseline pulse wave amplitude (P < 0.001) were independent predictors of ezFMD.
Conclusions:
These findings suggest that measurement of ezFMD, a novel noninvasive and simple method, may be useful for determination of vascular diameter response to reactive hyperemia. Since ezFMD is automatically measured by a device with an oscillometric method, measurement of ezFMD is easier and less biased than that of conventional FMD.