The IAEA/WHO TL dosimetry service for radiotherapy centres 1969-87
H Svensson1, G P Hanson, K Zsdánszky
1International Atomic Energy Agency, Vienna, Austria.
Acta Oncologica (Stockholm, Sweden)
|January 1, 1990
Summary
The International Atomic Energy Agency
Area of Science:
- Medical Physics
- Radiation Dosimetry
Background:
- The International Atomic Energy Agency (IAEA) initiated a postal dosimetry service in 1966.
- The World Health Organization (WHO) joined this initiative in 1968.
- Lithium fluoride (LiF) thermoluminescent dosimeters (TLD) were utilized.
Purpose of the Study:
- To ensure traceability to primary standards of absorbed dose.
- To assess the accuracy of radiotherapy dosimetry across numerous centers globally.
- To evaluate the effectiveness of the IAEA's postal dosimetry program.
Main Methods:
- 2,000 dosimeter intercomparisons were conducted over several decades.
- Dosimeters were sent to primary standards laboratories, including the 'Bureau International des Poids et Measures'.
- Thermoluminescent dosimeters (TLD) were used for measurements from radiotherapy centers in 89 countries, focusing on 60Co machines.
Main Results:
- The IAEA's dosimetry service demonstrated high accuracy, with a mean deviation of -0.26% ± 0.18% for traceability checks.
- Radiotherapy centers showed a mean deviation of -0.25% with a 6.7% SD for single results.
- Approximately 5% of centers occasionally exceeded 30% deviation, but accuracy improved with repeated participation.
Conclusions:
- The IAEA's postal dosimetry service effectively ensures traceability and improves radiotherapy dosimetry accuracy.
- Repeated participation in the TLD service significantly enhances the accuracy of dosimetry departments.
- The program has successfully involved around 700 radiotherapy centers worldwide, contributing to radiation safety.