Summary
Remote magnetic navigation (RMN) combined with the Carto 3 system effectively treats atrial fibrillation (AF) ablation while significantly minimizing fluoroscopic X-ray exposure. This approach offers a promising alternative for AF management with reduced radiation risks.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Technology
Background:
- Catheter ablation for atrial fibrillation (AF) primarily uses pulmonary vein isolation (PVI).
- PVI procedures are associated with substantial patient and operator X-ray exposure.
- Limited data exist on combining remote magnetic navigation (RMN) with the Carto 3 system for AF ablation.
Purpose of the Study:
- To evaluate the efficacy of RMN and Carto 3 system in AF ablation.
- To quantify the reduction in fluoroscopic exposure during AF ablation using this combined technology.
Main Methods:
- Eighty-one patients with symptomatic drug-refractory AF underwent RMN-guided PVI using the Carto 3 system.
- Radiofrequency ablation utilized a fixed power setting (≤ 35W).
- Primary endpoint was wide-area circumferential PVI; secondary endpoints included procedural outcomes and freedom from atrial tachycardia/AF.
Main Results:
- The study included 81 patients (mean age 60±9 years; 20% women).
- Median procedure time was 3.5±1 hours with a median X-ray exposure time of 13±7 minutes.
- After a median follow-up of 15±6 months, 71% of patients were symptom-free after one procedure, and 84% after two.
Conclusions:
- RMN with irrigated catheters and the Carto 3 system is effective for AF ablation.