Laying-up of sterile instruments in the operating theatre: equal or superior protection by using a horizontal
A A L Traversari1, C A Goedhart, E Dusseldorp
1TNO Dutch Centre for Health Assets, Soesterberg, The Netherlands.
Background:
A system for the preparation of sterilized instruments with unidirectional horizontal air flow (UDHF) has several advantages over a unidirectional down flow system (UDDF). The advantages are based on the installation of the system being more flexible and easier to use, no cooling of the air flow being necessary and less air being needed for circulation, resulting in reduced energy use.
Objectives:
The objective of this study was to determine whether a system with UDHF performs equal or superior to a system with UDDF in terms of prevention of contamination of the air (the presence of particles and micro-organisms) during the laying-up process.
Methods:
The degree of protection (DP) offered by two UDHF system variants and two UDDF system variants was determined for several static set-ups and a dynamic simulation of the process. In addition to determining the level of protection for several categories of particle size, colony-forming units (CFU) were also measured during process simulations.
Findings:
When maximum protection (no particles present) is considered, the UDHF systems performed significantly better than the UDDF systems for particles ≥2.5μm. When particles were present, there was no significant difference between systems for particles ≥0.3 and ≥0.5μm. However, the performance of the UDHF system was superior to that of the UDDF system (DP) for particles ≥1.0μm representing the bacteria-carrying particles. During the process measurements, no CFU were found with the UDDF system in 64% of the measurements, compared with 90% for the UDHF system (P = 0.012).
Conclusions:
The UDHF system offers equal or superior protection to the UDDF system against contamination of the clean area within which the laying up takes place. Despite our finding that the differences did not always reach statistical significance (due to low background concentrations), there is a clear trend, from the small-sized particles (≥1.0μm) up to the largest sizes considered, including bacteria-carrying particles, that demonstrates the superiority of the horizontal flow system. The UDHF system offers a more robust solution than the UDDF system, provided that good work instructions are given and the height of the table and height of the plenum are properly adjusted.
Related Concept Videos
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Asepsis
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...