Related Experiment Video
Updated: May 8, 2026

Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
Reproducibility of speckle-tracking-based strain measures of left ventricular function in a community-based study
Susan Cheng1, Martin G Larson, Elizabeth L McCabe
1Framingham Heart Study, Framingham, Massachusetts; Division of Cardiovascular Medicine, Department of Medicine, Brigham and Women's Hospital, Boston, Massachusetts.
Background:
The reproducibility of echocardiographic measurements of myocardial strain, performed in a community-based setting, has not been reported previously.
Methods:
The reproducibility of left ventricular strain measurements was examined in two samples of 20 participants each from the Offspring Cohort of the Framingham Heart Study (mean age, 63 ± 9 years; 59% women). Two-dimensional speckle-tracking-based measurements of global peak left ventricular strain in systole were performed in the apical four-chamber, apical two-chamber, and midventricular parasternal short-axis views.
Results:
Interobserver intraclass correlation coefficients were ≥0.84 for all global strain measurements, with average coefficients of variation of ≤4% for global longitudinal and circumferential strain and <8% for global transverse and radial strain. For left ventricular strain measurements performed in each of the three views, intraobserver intraclass correlation coefficients were ≥0.91 among time points spanning a total 8-month period. The average coefficients of variation were <6% for global longitudinal and circumferential strain and <9% for global transverse and radial strain. Interobserver and intraobserver reproducibility findings were similar in analyses adjusting for frame rate.
Conclusions:
Excellent reproducibility of global longitudinal and circumferential strain measurements and very good reproducibility of global transverse and radial strain measurements were observed. Taken together, these findings demonstrate the reproducibility of performing echocardiographic strain measurements in a large, epidemiologic community-based setting.
More Related Videos
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
07:21Magnetic Resonance Derived Myocardial Strain Assessment Using Feature Tracking