Evaluation and subsequent minimization of matrix effects caused by phospholipids in LC-MS analysis of biological
Franziska Janusch1, Lena Kalthoff, Gerd Hamscher
1Justus Liebig University Giessen, Institute of Food Chemistry & Food Biotechnology, Heinrich-Buff-Ring 58, 35392 Giessen, Germany.
Background:
The influence of matrix effects in LC-MS/MS analysis of biological samples can be enormous and has to be evaluated during method development. Phospholipids, which are present in considerable quantities in biological fluids, are supposed to cause matrix effects when co-eluting with analytes. Therefore, the reduction of phospholipids should lead to the minimization of matrix effects. METHODOLOGY & RESULTS: Here, a polymeric reversed-phase (PRP) SPE cartridge was compared with a combination of mixed-mode-anion-exchange (MAX) and mixed-mode-cation-exchange (MCX) SPE cartridges regarding elimination of matrix effects during sample clean-up. For evaluation of matrix effects post-column infusion experiments were performed. Phospholipid amount in the sample extract and matrix effects are enhanced using PRP in contrast to the combination of MAX/MCX.
Conclusion:
For an efficient elimination of phospholipids during sample preparation and to improve method accuracy and precision it is advisable to use a combination of MAX/MCX SPE cartridges.
More Related Videos
10:37Deep Proteome Profiling by Isobaric Labeling, Extensive Liquid Chromatography, Mass Spectrometry, and Software-assisted Quantification
Published on: November 15, 2017
07:33Lipid Vesicle-mediated Affinity Chromatography using Magnetic Activated Cell Sorting (LIMACS): a Novel Method to Analyze Protein-lipid Interaction
Published on: April 26, 2011