Comparison of stroke volume and fluid responsiveness measurements in commonly used technologies for goal-directed
Simon J Davies1, Simran Minhas, R Jonathan T Wilson
1Department of Anaesthesia, York Hospital, NHS Foundation Trust, Wigginton Rd., York YO31 8HE, UK.
Study Objective:
To compare stroke volume (SV) and preload responsiveness measurements from different technologies with the esophageal Doppler monitor (EDM).
Design:
Prospective measurement study.
Setting:
Operating room.
Patients:
20 ASA physical status 3 patients undergoing vascular, major urological, and bariatric surgery.
Interventions:
Subjects received fluids using a standard Doppler protocol of 250 mL of colloid administered until SV no longer increased by >10%, and again when the measured SV decreased by 10%.
Measurements:
Simultaneous readings of SV, stroke volume variation (SVV) and pulse pressure variation (PPV) from the LiDCOrapid, and SVV from the FloTrac/Vigileo were compared with EDM measurements. The pleth variability index (PVI) also was recorded.
Main Results:
No correlation was seen in percentage SV change as measured by either the LiDCOrapid (r=0.05, P=0.616) or FloTrac (r=0.09, P= 0.363) systems compared with the EDM. Correlation was present between the LiDCOrapid and FloTrac (r=0.515, P<0.0001). Percentage error compared with the EDM was 81% for the FloTrac and 90% for the LiDCOrapid. SVV as measured by LiDCOrapid differed for fluid responders and nonresponders (10% vs 7%; P=0.021). Receiver operator curve analysis to predict a 10% increase in SV from the measured variables showed an area under the curve of 0.57 (95% CI 0.43-0.72) for SVV(FloTrac), 0.64 (95% CI 0.52-0.78) for SVV(LiDCO), 0.61 (95% CI 0.46 -0.76) for PPV, and 0.59 (95% CI 0.46 -0.71) for PVI.
Conclusions:
Stroke volume as measured by the FloTrac and LiDCOrapid systems does not correlate with the esphageal Doppler, has poor concordance, and a clinically unacceptable percentage error. The predictive value of the fluid responsiveness parameters is low, with only SVV measured by the LiDCOrapid having clinical utility.
Related Concept Videos
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...
Measurement of Blood Pressure