Comparison of custom to standard TKA instrumentation with computed tomography
Vincent Y Ng1, Lindsay Arnott, Jia Li
1Department of Orthopaedics, The Wexner Medical Center, The Ohio State University, 725 Prior Hall, 376 West 10th Avenue, Columbus, OH, 43210, USA, vincent_y_ng@yahoo.com.
Purpose:
There is conflicting evidence whether custom instrumentation for total knee arthroplasty (TKA) improves component position compared to standard instrumentation. Studies have relied on long-limb radiographs limited to two-dimensional (2D) analysis and subjected to rotational inaccuracy. We used postoperative computed tomography (CT) to evaluate preoperative three-dimensional templating and CI to facilitate accurate and efficient implantation of TKA femoral and tibial components.
Methods:
We prospectively evaluated a single-surgeon cohort of 78 TKA patients (51 custom, 27 standard) with postoperative CT scans using 3D reconstruction and contour-matching technology to preoperative imaging. Component alignment was measured in coronal, sagittal and axial planes.
Results:
Preoperative templating for custom instrumentation was 87 and 79 % accurate for femoral and tibial component size. All custom components were within 1 size except for the tibial component in one patient (2 sizes). Tourniquet time was 5 min longer for custom (30 min) than standard (25 min). In no case was custom instrumentation aborted in favour of standard instrumentation nor was original alignment of custom instrumentation required to be adjusted intraoperatively. There were more outliers greater than 2° from intended alignment with standard instrumentation than custom for both components in all three planes. Custom instrumentation was more accurate in component position for tibial coronal alignment (custom: 1.5° ± 1.2°; standard: 3° ± 1.9°; p = 0.0001) and both tibial (custom: 1.4° ± 1.1°; standard: 16.9° ± 6.8°; p < 0.0001) and femoral (custom: 1.2° ± 0.9°; standard: 3.1° ± 2.1°; p < 0.0001) rotational alignment, and was similar to standard instrumentation in other measurements.
Conclusions:
When evaluated with CT, custom instrumentation performs similar or better to standard instrumentation in component alignment and accurately templates component size. Tourniquet time was mildly increased for custom compared to standard.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
13:16Computed Tomography and Optical Imaging of Osteogenesis-angiogenesis Coupling to Assess Integration of Cranial Bone Autografts and Allografts
Published on: December 22, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...