Multidetector CT radiation dose optimisation in adults: short- and long-term effects of a clinical audit
Denis Tack1, Andreas Jahnen, Sarah Kohler
1Department of Radiology, EpiCURA Hospital, Clinique Louis Caty, Rue Louis Caty 136, B 7331, Baudour, Belgium, denis.tack@skynet.be.
Objective:
To report short- and long-term effects of an audit process intended to optimise the radiation dose from multidetector row computed tomography (MDCT).
Methods:
A survey of radiation dose from all eight MDCT departments in the state of Luxembourg performed in 2007 served as baseline, and involved the most frequently imaged regions (head, sinus, cervical spine, thorax, abdomen, and lumbar spine). CT dose index volume (CTDIvol), dose-length product per acquisition (DLP/acq), and DLP per examination (DLP/exa) were recorded, and their mean, median, 25th and 75th percentiles compared. In 2008, an audit conducted in each department helped to optimise doses. In 2009 and 2010, two further surveys evaluated the audit's impact on the dose delivered.
Results:
Between 2007 and 2009, DLP/exa significantly decreased by 32-69 % for all regions (P < 0.001) except the lumbar spine (5 %, P = 0.455). Between 2009 and 2010, DLP/exa significantly decreased by 13-18 % for sinus, cervical and lumbar spine (P ranging from 0.016 to less than 0.001). Between 2007 and 2010, DLP/exa significantly decreased for all regions (18-75 %, P < 0.001). Collective dose decreased by 30 % and the 75th percentile (diagnostic reference level, DRL) by 20-78 %.
Conclusions:
The audit process resulted in long-lasting dose reduction, with DRLs reduced by 20-78 %, mean DLP/examination by 18-75 %, and collective dose by 30 %.
Key Points:
• External support through clinical audit may optimise default parameters of routine CT. • Reduction of 75th percentiles used as reference diagnostic levels is 18-75 %. • The effect of this audit is sustainable over time. • Dose savings through optimisation can be added to those achievable through CT.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT