Related Experiment Video
Updated: May 8, 2026

Contrast Enhanced Vessel Imaging using MicroCT
Published on: January 27, 2011
Energy-discriminative performance of a spectral micro-CT system
Peng He1, Hengyong Yu, James Bennett
1The Key Lab of Optoelectronic Technology and Systems of the Education Ministry of China, Chongqing University, Chongqing, China.
Abstract:
Experiments were performed to evaluate the energy-discriminative performance of a spectral (multi-energy) micro-CT system. The system, designed by MARS (Medipix All Resolution System) Bio-Imaging Ltd. (Christchurch, New Zealand), employs a photon-counting energy-discriminative detector technology developed by CERN (European Organization for Nuclear Research). We used the K-edge attenuation characteristics of some known materials to calibrate the detector's photon energy discrimination. For tomographic analysis, we used the compressed sensing (CS) based ordered-subset simultaneous algebraic reconstruction techniques (OS-SART) to reconstruct sample images, which is effective to reduce noise and suppress artifacts. Unlike conventional CT, the principal component analysis (PCA) method can be applied to extract and quantify additional attenuation information from a spectral CT dataset. Our results show that the spectral CT has a good energy-discriminative performance and provides more attenuation information than the conventional CT.